  • 摘要: 试验探讨了DC Arc直流电弧光谱仪测定纯钽及钽制品中Fe元素的条件,并建立了测定方法。通过试验确定了Fe元素的分析线波长、积分时间及背景点位置。按照实验方法测得样品中Fe元素的加标回收率在92.73%—107.27%之间,含量测定的相对标准偏差(RSD,n=10)为4.8%,Fe元素的检出限为1.41μg/g。本法对标准样品的测定结果与GE-340光栅光谱法、DV-5直流电弧光谱法的测定结果基本一致。

     

    Abstract: Determination conditions for Fe in pure tantalum and tantalum products were discussed and the determination method established. The wavelengths,integral time and point location in the background of Fe were determined by tests. According to the experiment,the standard addition recovery is between 92.73% and 107.27%,the relative standard deviation(RSD,n = 10) of Fe content is 4.8%,and the Fe detection limit is 1.41 μg/g. The determination results presented by this method are basically in accordance with those by GE-340 grating spectrum method and DV-5 dc arc spectrum method.

     

/